Operating limits are set, predefined tabular limits known as STANDARD, CIVIL and MAX these
can be selected through the SETUP menu (see Section 7.3.3, "SETUP, [LIMITS]," on page 83).
Operators may also configure the display to aeronautical or pressure units but should be
aware that when units of pressure are selected, wider full-scale pressure limits will be
enabled for some parameters.

1.3 Operating Limits

The following sets of operating limits are supplied with the ADTS 405 Mk2.
Parameter
Min Alt
Max Alt
Min CAS
Max CAS
Min Ps
Max Ps
Min Qc
Max Qc
Max Mach
Max ROC
Max Rate CAS
Max Rate Ps
Max Rate Qc
ARINC Limits
Alt Correction
2 | ADTS 405 R/F Mk2 Instruction Manual-English
Table 1-1: Operating Limits - 850 kts Range
Max Limits
-3,000 ft
105,000 ft
-100 kts
850 kts
3.0 mbar
1355.00 mbar
-1,352.00 mbar
1,700.00 mbar
12.35
100,000 ft/min
2,000 kts/min
10,000.00 mbar/min
10,000.00 mbar/min
OFF
0 ft
Standard Limits
-2,000 ft
80,000 ft
0.0 kts
850 kts
27.615 mbar
1088.657 mbar
0.0 mbar
1688.00 mbar
2.5
9,000 ft/min
600 kts/min
200.00 mbar/min
200.00 mbar/min
OFF
0 ft
© 2017 General Electric Company – All Rights Reserved.
Civil Limits
-1,000 ft
50,000 ft
0 kts
450 kts
115.972 mbar
1050.406 mbar
0.0 mbar
368.01 mbar
1
6,000 ft/min
600 kts/min
100.00 mbar/min
100.00 mbar/min
OFF
0 ft