sauter FL Instruction Manual page 20

Digital force gauge
Hide thumbs Also See for FL:
Table of Contents

Advertisement

6.1. Zero Compression
• Set up the measuring device and all necessary accessories for performing the
compression calibration
• apply a force of 100% of the meter's power to the weight
• Remove the load, leave the device standing
• as soon as the display becomes stable, press the ZERO key to save the zero
display
6.2. maximum compression
• apply the compression force of 100% of the gauge power
Calibrate the compression gain with the UP and DOWN arrow keys and
RIGHT and LEFT until the meter displays the maximum compression result
• touch the ENTER key to save the results for the entire compression scale
7. After successful calibration, remove all accessories from the meter and place the
meter on a flat surface with its backside facing up.
Select the voltage increase menu window, press the ZERO key to obtain the
primary calibration offset, date and time and reset the overload counter.
20
FL-BA-e-2020