Installing Intuvo 9000 GC Column
To Install a Column Guard or Jumper Chip
3 Wait until the GC reaches the ready state indicating the components are
4 If you are using hydrogen or other flammable gas as a carrier gas or for the
5 If installed, remove the ALS injector from the inlet.
6 Remove the inlet cover. (See
GC inlet cover
removed
Compression bolt
access plate
Figure 11.
7 Open the GC front door.
8 Open the bus door and remove it by lifting the door vertically off its hinge
9 Pull the compression bolt access plate out to allow the torque driver to
7000/7010 Series TQ GC/MS Operating Manual
cooled below 40 °C and the instrument is vented before continuing the steps
in this procedure.
JetClean system, close the manual gas supply valve to the instrument before
turning off the MS power.
GC inlet cover and compression bolt access plate
pins.
access the Guard chip compression bolt. (See
Figure
11.)
Figure 12
on page 60.)
59