Installing 8890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using the Self-Tightening Column Nut
15 Gently check the alignment of the ion source and the interface tip seal. When
16 You can align the ion source and interface tip seal by wiggling the side plate
17 Close the front analyzer chamber. (See
7000/7010 Series TQ GC/MS Operating Manual
the ion source is aligned correctly, the front analyzer chamber can be closed
all the way with no resistance except the spring tension from the interface tip
seal.
on its hinge. If the door still will not close, contact your Agilent Technologies
service representative.
Chamber"
on page 195.)
"To Close the Front Analyzer
41