Operating in CI Mode
To Perform a CI Autotune
8 Select the Tune from default settings check box if you are restarting the
9 Select the Save tune file when done check box to save the new tune
10 Select the Print autotune report check box to automatically print a tune
11 Click Autotune to start the autotune. The Status line displays the current step
12 Review the Tune Report. If the results are acceptable and you did not select
7000/7010 Series TQ GC/MS Operating Manual
system after a system vent, major servicing, or a power outage. If you clear
the Tune from default settings box, the Autotune process starts using
previous tune values.
parameters generated by the autotune. Do not select this item if you want to
review the autotune report before saving the newly generated tune
parameters.
report.
in the autotune process, and the plot of the tuned parameter for that step is
shown in the top graph. If specified, at the completion of the autotune a Tune
Report is printed.
To stop the autotune before it completes the autotune parameter selection,
click Abort Autotune. The parameters from the last successful autotune are
used.
the Save tune file when done check box, save the autotune by clicking the
Files and Report tab, then click Save.