Agilent Technologies 7000 Series Operating Manual page 55

Hide thumbs Also See for 7000 Series:
Table of Contents

Advertisement

3
Installing Intuvo 9000 GC Column
To Replace an Intuvo GC Column
Click and run
connectors
Compression
bolts
Column clamps
Retainer ring
SmartID key
Figure 10.
8 Unplug the column SmartID key from the lower USB port.
9 Using the Intuvo torque driver, remove the two compression bolts that seal
10 Remove and store the column for later use in accordance with the column
11 Replace the gasket with a new gasket rated for the maximum column
7000/7010 Series TQ GC/MS Operating Manual
9000 GC column and related parts
the column's click and run connectors to the bus and store for later use.
manufacturer's recommendations.
temperature expected with your method. (See
GC Gasket"
on page 57.)
"To Replace an Intuvo 9000
55