General Maintenance
To Remove the EI XTR Source
3 Disconnect wiring from the EI XTR. Do not bend the wires any more than
4 Remove the thumbscrews that hold the ion source in place. (See
5 Pull the ion source out of the source radiator.
Ceramic source board
Ion source
Thumbscrews
Source heater and temperature sensor wires
Source radiator
Figure 50.
Removing the EI XTR source
7000/7010 Series TQ GC/MS Operating Manual
necessary. (See
"To Connect or Disconnect Wiring to the EI XTR Source"
page 164.)