Michell Instruments XTP601 User Manual page 44

Process oxygen analyzer
Hide thumbs Also See for XTP601:
Table of Contents

Advertisement

APPENDIX A
Appendix A
Technical Specifi cations
Performance
Measurement Technology
Gas
Measurement Range
Display Resolution
Display Type
Accuracy (excluding suppressed
zero ranges)
Accuracy (for suppressed zero
ranges 20/80/90-100%)
Response Time (T90) with HSR
enabled
Repeatability
Linearity
Zero Stability
Span Stability
Sample Flow Rate
Sample Pressure
Sample Temperature
Sample Cell Temperature
Background Gas
Electrical Input/Output
Analog Inputs
Analog Outputs
Output Ranges
Alarms
Datalogging
Digital Communications
Power Supply
Cable Entries
36
Thermo-paramagnetic oxygen sensor
Process and non-condensing sample with particles <3μm
Select from 0-0.5% up to 0-50% and 20-100% up to 90-100%
0.01% (0.1% for suppressed zero ranges)
Backlit LCD
< ±1% of span or ±0.02% O
< +/-1% of Span or 0.2% O
< 15 seconds
±0.2% of span or 0.01% O
±0.5% of span or 0.05% O
±0.25% of span per month
±0.25% of span per month
GP1 and GP3: 100 to 500 ml/min (0.2 to 1.06 scfh)
EX1, EX3 and GP2: 270 to 330 ml/min (0.6 to 0.7 scfh)
0.75 to 5 BarA (11 to 72.5 psiA)
5 to 55°C (+41 to +122°F) max
(+45°C (+113°F) for standard cell)
Standard +50°C (+113°F)
(Optional +55°C and +60°C (+131°F and +140°F) for ranges
0-0.5% up to 0-50% only.)
Analyzer is calibrated in the background gas of the process.
Standard gases are: N2, CO2, CH4, H2 & Biogas. Others available
on request.
2 off 4-20 mA inputs
One for an external sensor that can be displayed on the screen
One to act as an active compensation for the process conditions
2 off 4-20 mA outputs (powered with 24V excitation voltage)
Primary range is set to the calibrated range of the instrument
The second is user selectable 0-100%
2 off single pole changeover (SPCO) relays for O
(250 V, 5 A max)
The user can use the digital communications to log the output
from the analyzer
The unit will store 40 alarm points and the min/max O
concentrations with date and time stamp
Modbus RTU over RS485 Protocol
24 V DC; 1.5 A max
Analyzer is supplied with 3 x M20 cable entries. Cable glands,
conduit entries and blanking plugs are available as accessories.
XTP601 User's Manual
whichever is greater
2
whichever is greater
2
whichever is greater
2
whichever is greater
2
2
97313 Issue 5, May 2018
concentration
2

Advertisement