Before the use

Setting up and connecting the centrifuge

The setting up and connecting of the
Cryofuge 5500i GMP may be imple-
mented exclusively by trained special-
ist personnel only!

The correct installation locations

The utilization of the centrifuge has to be implemented
in interior compartments. The installation locations
must meet these requirements:
A security area of at least 30 cm must be kept
around the centrifuge. No dangerous materials may
be located there during the centrifugation.
The floor must be stable and resonance-free.
The centrifuge must be protected against heat and
strong solar radiation.
UV radiation reduce the durability of
plastics.
Place the centrifuge, as well as ro-
tors and accessories of plastic, so
that they are not exposed to direct
solar radiation!

Transporting centrifuge

The centrifuge can be damaged
during
transport
impacts!
Transport the centrifuge in the
upright position, without rotor and
only in the corresponding special
packing.
Ensure sufficient protection!
Before the use
through
shock
13