Operation

Stopping the centrifuge

At limited run time

Normally the run time has been set manually and all
you have to do is wait until the centrifuge terminates
the run automatically at the end of the set time.
As soon as the speed reaches zero, the display reads
"end". You can now open the centrifuge by pressing
the "open lid" key
material.
If the lid has not been fully lifted out of the lid lock, the
message "lift lid" appears (manual lift of the lid is nec-
essary).
You can manually stop the centrifuge at any time by
actuating the "stop" key
At this point the remaining run time is displayed.

At continuous operation

If you have chosen continuous operation, you must
stop the centrifuge manually. Press the "stop" key
in the control panel. The centrifuge will be deceler-
ated with the preset deceleration profile.
The display reads "end", and you can open the lid by
pressing the "open lid" key
fuged material.
56
and remove the centrifuged
.
and remove the centri-

Temperature control during standby

The exact regulation will be active after identification
of the rotor. This is the case after a centrifugation run
exceeding 300 rpm. At standby the speed display
reads "end".
If the rotor has not been identified (lid has been
closed and the "start" key
ated, speed display shows "0" with flashing point), the
temperature regulation inhibits the freezing of the
sample only. An exact regulation cannot be guaran-
teed.