Dräger Infinity Delta Series Supplement Manual page 99

Table of Contents

Advertisement

Static cuff accuracy
Calibration range
Continuous Non-Invasive Arterial Blood Pressure via CNAP Pod
Parameter label
Systolic
Diastolic
Mean
Units
Measuring range
Systolic
Diastolic
Mean
HR range (for CNAP measurements)
Accuracy
Display resolution
Inflation pressure
Default
Minimum
Maximum
Over pressure limit
Pressure
Response time
Deflation time
Supplement – Infinity
®
±3 mmHg (± 0.4 kPa)
Adult and Pediatric:10 to 260 mmHg ±3 mmHg (1.3 to 34.7 kPa ±
0.4 kPa)
Neonatal: 10 to 150 mmHg ±3 mmHg (1.3 to 20.0 kPa ± 0.4 kPa)
Delta Series Software VF10.1
Supplement for Infinity Delta Series Software
CNAP
CNAP-S
CNAP-D
CNAP-M
NOTE: Only values for systolic, diastolic, and
mean are displayed in the p-box. Labels are not
displayed.
mmHg (kPa)
40 to 250 mmHg (5.3 to 33.3 kPa)
30 to 210 mmHg (4 to 28 kPa)
35 to 230 mmHg (4.66 to 30.7 kPa)
20 to 200 bpm
± 5 mmHg (0.6 kPa)
1 mmHg (0.1 kPa)
120 mmHg (16 kPa)
30 mmHg (4 kPa)
300 mmHg (40 kPa)
300 mmHg (40 kPa) + 10%
≤ 3 s
≤ 15 s (to reach < 15 mmHg)
99