Supplement for Infinity Delta Series Software

NBP Setup

NOTE
The accuracy of the oscillometric blood pressure signal can decrease (up to loss of measurement)
under the following conditions:
– Weak pulses
– Irregular pulses
– Patient movement artifacts
– Tremor artifacts
– Respiratory artifacts
– Pulses generated from a ventricular assist
device
These conditions may prevent the monitor from performing a measurement within the maximum
allowable measurement time.
To connect the hose
 Push the hose end fitted with the plastic collar firmly into the connector on the left side of the monitor
(see figure below).
NBP hose connector
Delta/Delta XL
Venous Stasis
Available in software levels equal to or greater than VF9.1, and supports venous stasis in adult mode only
(with adult-based inflation pressures).
28
NBP hose connector
Kappa
Supplement – Infinity
®
Delta Series Software VF10.1