Status Messages - Dräger Infinity Delta Series Supplement Manual

Table of Contents

Advertisement

Status Messages

Message
Possible Cause
NBP Cannot
The pulse profile is too poor to make a
Measure
reliable measurement (usually due to
persistent motion artifact).
NBP
A measurement was aborted (usually
Measurement
due to motion artifact) because it lasted
Timeout
longer than two minutes (adult or
pediatric) or 90 seconds (neonatal).
Supplement – Infinity
®
Delta Series Software VF10.1
Supplement for Infinity Delta Series Software
Suggested Action
• Check the patient and treat, if
necessary.
• Move the cuff to a limb with less
movement.
• Restart measurement. If the
message persists, contact the
hospital's technical personnel or
DrägerService.
• Take a manual measurement if the
condition persists.
• Repeat the measurement.
• Take a manual measurement if the
condition persists.
29