ECG Setup Menu

Menu selection
Description
The ECG Options submenu
This submenu allows you to configure the following functions:
Pacer Detection
Determines the monitor's ability to
identify pacemaker pulse. Allows user
to enable/disable pacer detection or
choose more advanced Fusion
selection.
CAUTION: Fusion mode pacer
detection is not intended for use with
large-signal, unipolar pacemakers. It is
intended for use only with bipolar
pacemakers.
Pacer Detection
Observe the following:
(continued)
– Select Fusion mode only in
– Before selecting Fusion mode, be
– Do not use Fusion mode if you are
Supplement – Infinity
®
situations where it becomes
necessary to suppress repeated
false asystole and/or false low heart
rate alarms.
certain that the patient has a bipolar
pacemaker (external or implanted)
and that it is accurately
programmed as appropriate for that
patient.
uncertain as to what type of
pacemaker is being used.
Delta Series Software VF10.1
Supplement for Infinity Delta Series Software
Available Settings
In basic mode:
– Select ON to enable pacemaker
detection.
– Select OFF (default) to disable
pacemaker detection.
In advanced mode:
– Select ON to enable pacemaker
detection.
– Select OFF (default) to disable
pacemaker detection.
– Select Fusion to enable pacer
detection, but minimize pacer tail
rejection to reduce missed detection of
pseudo-fused paced beats, which
results in false asystole alarms.
23