Dräger Infinity Delta Series Supplement Manual page 48

Table of Contents

Advertisement

Supplement for Infinity Delta Series Software
Menu Item
Description
Scale
Sets the etCO2* waveform scale.
Resp. Sweep
Sets the respiratory waveform sweep speed on
Speed
the screen's display.
Agent Display
Displays a separate agent parameter box.
Ghosted if the monitor is displaying combined
MultiGas parameter box.
Pressure Comp. Sets the compensation for ambient atmospheric
pressure.This item is always set to auto and is
ghosted.
RRc* Apnea
Sets the time that the monitors waits before
Time
reporting a cessation of breathing as an apnea
event.
Apnea Archive
Stores and/or records an alarm event
automatically for apnea. You can later review
stored alarms on the Event Recall screen.
MultiGas Zero
Manually zeros the SCIO module (parameter
values disappear during zeroing).
Auto Zero Delay Delays automatic zeroing for 5 minutes for
uninterrupted monitoring.
WARNING: Delaying the auto zero may
impact accuracy.
NOTE: Gas sensors in the SCIO module are
automatically zeroed and calibrated against
room air. Parameter values disappear during
zeroing. One minute before automatic zeroing,
the monitor sounds an attention tone and
displays the message Auto zero in <1 minute.
etCO2* Alarm
Accesses etCO2* alarms in Alarm Limits table (see Chapter 5 of the main IFU)
WARNING
etCO2* alarms do not activate until the first breath is detected after turning on
the monitor or discharging a patient.
48
Settings
40, 60, 80, 100 mmHg (40 mmHg
default)
5, 8, 12, 16 kPa (5 kPa default)
5, 8, 12, 16% (5% default)
6.25, 12.5, 25, 50 mm/s
(6.25 mm/s default)
– ON
– OFF
– Auto
OFF, 10, 15, 20, 25, 30 s
OFF, 10, 15, 20, 25, 30 s
(Adult: OFF default)
(Pediatric/neonatal: 15 s default)
Not applicable
Not applicable
Supplement – Infinity
®
Delta Series Software VF10.1