UVP Minidizer Hybridization Oven
Operation
Caution: This equipment is not intended for interconnection with any other devices. Use of this equipment other than as
intended may create a safety hazard and/or malfunction.
Using the Hybridization Oven
1.
Place the unit on a level working surface and provide adequate room in front of the door to allow easy opening.
2.
Plug the female end of the power cord into the back of the unit and the male end into a surge-protected outlet.
3.
To turn on the main power supply, press the Power switch toward "ON" on the right side of the control panel.
4.
To turn on the motor control to rotate the carousel, press the motor control power switch to the "ON" position.
Temperature Calibration Controls
When the power is turned on, the software
revision number flashes for several
seconds. The display will then show
internal surface temperature (not air
temperature) for a few seconds. The
display will then revert to the factory
preset temperature of 30°C.
Altering the Temperature
Setpoint
The current setpoint value can be altered
using the "UP" and "DOWN" buttons while
the setpoint is being displayed. To change the setpoint from normal
mode, proceed as follows:
1.
Press then release either the "UP" or the "DOWN" button.
The LCD will flash to indicate that the displayed value is the current setpoint. If either button is held
down for more than several seconds, setpoint value will increase or decrease incrementally.
2.
When the desired setpoint is reached, wait approximately five seconds without pressing either button;
the display will revert to normal mode showing the actual temperature. The new setpoint becomes
effective. When the unit is turned off and turned on again, the last used setpoint will be displayed.
3.
Allow sufficient time for warm-up prior to hybridization procedures.
Temperature Calibration
The hybridization oven is calibrated at the Analytik Jena factory. AJ recommends temperature recalibration be
performed at the AJ factory as recalibration by the user may void the warranty. To return the unit to AJ, first
obtain a RGA (Returned Goods Authorization) number from AJ.
Call AJ's customer service department in Upland, California at (909) 946-3197 or Cambridge, UK at +44(0)1223-
420022 to obtain an RGA number or to inquire about temperature calibration.
Service Procedures
Care and Cleaning
NOTE: ALWAYS UNPLUG THE UNIT FROM THE ELECTRICAL SUPPLY BEFORE CLEANING OR DRYING THE UNIT!
Temperature
Display: Press the
up and down
arrows to set the
temperature
Power: Press the
power switch to
the "I" position
operate the
temperature
Motor: Press the
power switch to
the "I" position
rotate the carousel
6
81-0169-02 Rev J