For Your Safety:

PLEASE READ THIS MANUAL IN ITS ENTIRETY BEFORE ATTEMPTING INSTALLATION
OR OPERATION!
understanding its features and functions may result in unsafe conditions.
Always use protective eye wear and observe proper safety procedures when working with pressurized
gases.
Always assure the pressure of gas entering the Model 5100E sensor is 1-10 psig. Not to exceed 20 psig
Always calibrate the Model 5100E at an equivalent pressure and flow rate to the measured gas.
Maximum pressure should not exceed 10 psig.
Always allow the Model 5100E to cool down before attempting to access the sensor.
Ensure the Model 5100E has been properly calibrated before use.
Never expose the Model 5100E analyzer chassis to water, high humidity or moisture. The analyzer
chassis is not watertight.
Never expose the Model 5100E to flame or high temperatures.
Never expose the Model 5100E analyzer to flammable gases or vapors. The unit is not rated Explosion
Proof, or Intrinsically Safe.
Never expose the Model 5100E directly to unregulated gas lines, cylinder gas. High gas pressures may
cause the sensor components to rupture.
Ensure the analyzer unit is mounted in an area of free airflow to prevent the chassis from exceeding the
operating temperature specifications. Do not mount the analyzer against hot surfaces. Do not block
the ventilation louver on the analyzer chassis.
Attempting to operate the Model 5100E without fully