2.1.2 Step 2 –Install the Remote Sensor
The Model 5100E is supplied with a Model ZR-500E oxygen sensor, and sensor flow-through head for
connection to a sampled process gas stream, and a sensor interface cable. The Model 5100E can also be
supplied with a Neutronics Inc. process Sampling system, built-to-application. NOTE: For instructions on
the installation of a Neutronics Inc. Process Sampling System, please refer to the sampling system
equipment manual.
2.1.2.1
Flow-through Head
Surface-mount the flow through head horizontally (as shown in figure 9) or vertically on a stationary panel.
The sensor flow-through head is 1.25" diameter stainless steel, and is machined to accommodate two # 6-
32 machine-type mounting screws (1-inch on center). Be careful not to over tighten the mounting screws.
Allow sufficient space to screw the Model ZR-500E oxygen sensor into the top threaded port of the flow
through head, and for the sample lines and sample inlet and exhaust fittings.
7-PIN PLUG, DIN
THREADED CONNECTOR - 3/4-16
GASKET
1/8" FNPT -
SAMPLE INLET
Figure 9 – Sensor mounting detail
2.1.2.2
Sample Inlet Port
Pneumatic connection to the measured process for sample extraction is made at either of the two
interchangeable 1/8" FNPT fitting around the side of the flow-through head. For connecting the flow-
through head to the measured process, use 1/8" or 1/4" rigid tubing, and 1/8" MNPT fittings of a material
compatible with process gas composition. Ensure that no grease, particulate, or solvent is present in the
tubing during installation. Use thread-tape to seal connections, and prevent gauling. Fix all sample tubing
and connectors.
A fixed calibration port may be implemented in the process sampling line by installing a 1/8" or 1/4" 3-way
manual ball valve into the sampling line as in figure-9. Use 1/8" or 1/4" rigid tubing and 1/8" MNPT fittings
of a material compatible with process gas composition. Ensure that no grease, particulate, or solvent is
present in the tubing during installation. Use thread-tape only to seal connections, and prevent galling. Fix
all sample tubing and connectors.
Manual Part Number:
C5-06-4900-08-0
SENSOR
O-RING TOP SEAL
SENSOR RECEPTACLE
1/8" FNPT -
SAMPLE
EXHAUST
MOUNTING SCREW
INSERT - #6-32
Revision Level: C
ECO 8029
MOUNTING HOLES, 2
PLACES, #6-32 THREAD
#36 (.106) DIA.
File Name: MN-A-0008
Installing the Analyzer
Page 2-3