The STANDARD ACCESS level commands
TYPED COMMAND
A
C
E
G
H
I
V
@
There are several data formats of the ASCII data dump available. They may be changed from the user
specified format (refer to the Software Setup Log that is supplied from the factory) to any format listed
below:
TYPED COMMAND
SSERFMT=0
SSERFMT=1
SSERFMT=2
SSERFMT=3
SSERFMT=4
SCALIBRATE=.XXXXXX
4.1.2.4.1 Disable RS 232 continuous OUTPUT ON REQUEST – SSERFMT=0
While this setting is active, the user must request information by pressing the desired key according to the
STANDARD ACCESS level commands chart.
4.1.2.4.2 Human Readable Data Format – SSERFMT=1
Human Readable data is presented in dynamic columns (columns appear only when data is present). It is
intended for most users, to aid in setup and maintenance of the unit. Column headings from left to right:
Mode • O
Concentration • Alarm-1 status • Alarm-1 status • list of Fault codes active.
2
4.1.2.4.3 Machine Data Format with NO Checksum
Machine format with NO checksum can be selected via the analyzer control panel, or through the RS-232
interface in Standard Access level. Machine format with NO checksum data is streamed in packets defined
by start/stop transmit bits. The order of data in each packet is as follows: Start Transmit • O
• Fault codes active • List of Fault Codes • Alarm-1 status • Alarm-1 status • End Transmit. For detailed
information on data formats, please contact the Neutronics Service Department.
Manual Part Number:
C5-06-4900-08-0
DESCRIPTION OF QUERIED FUNCTION
Short software version
Analyzer Model number
Sensor output in Percent Oxygen
Sensor output in Volts
Active Fault codes (32 bit codes in hex)
Active Fault code descriptions
Long software version
Unit Serial Number
DESCRIPTION OF QUERIED FUNCTION
Disables RS-232 continuous OUTPUT ON REQUEST
Enables RS-232 output in HUMAN READABLE format
Enables RS-232 output in MACHINE format w/o Checksum
Enables RS-232 output in MACHINE format w/Checksum
Enables RS-232 output in TAB DELIMITED (Excel®) format
The user can send a calibration value in decimal format where 100 %
Oxygen = 1. The number format entered must be 6-decimal places.
Revision Level: C
ECO 8029
File Name: MN-A-0008
System Setup
Concentration
2
Page 4-5