Bacharach Neutronics 5100E Operation Manual page 59

Oxygen analyzer / controller
Table of Contents

Advertisement

RANGE
MEASUREMENT
RANGE
ACCURACY OF
MEASUREMENT
RANGE
LDL
RESPONSE TIME
WARM UP TIME
HUMIDITY
OPERATING
TEMPERATURE
STORAGE
TEMPERATURE
SAMPLE PRESSURE
SAMPLE FLOW
POWER
MECHANICAL
WEIGHT
WARRANTY
**Measurement of oxygen >95% affects sensor life significantly, e.g. measuring 100% oxygen will potentially
reduce sensor life to less than 6 months.
concentrations.
The manufacturer reserves the right to amend these specifications as necessary.
Manual Part Number:
C5-06-4900-08-0
0–1 % • 0–10 %• 0–25 % • 0–50 % • 0–100 % • Low Auto (0–25 %)
• High Auto (10–100 %) • Full Auto (0–100 %)
0-25%, 0-100%**
+ .5% of measurement range
.05% @ 0-25% measurement range, .1% @ 0-100% measurement range
T
< 15 seconds
90
60-seconds for sensor warm-up, 35 minutes for temperature equilibrium
of the entire analyzer.
0-95 % non-condensing
Analyzer:
32-122° F (0–50° C)
Sensor:
32-122° F (0–50° C)
Analyzer:
23-149° F (-5–65° C)
Sensor:
23-149° F (-5–65° C)
15" Hg vacuum–7 PSIG (0.5–1.5 Bar)
0.5–1.5 LPM (1 LPM nominal) (1-3 cubic feet per hour)
VAC Unit:
90 – 264 VAC, 47 – 63 Hz, Single Phase, 13 Watts
VDC Unit:
11 – 18 VDC, 13 Watts
20 – 30 VDC, 13 Watts
Mains power is to be regulated to +/- 10%
Height 3.75"x Width 7.00" • NEMA 4 • IP66
Faceplate:
Panel Cut-out:
Height 2.91" x Width 6.20"
Height 2.81" x Width 5.98" x 3.60" Depth • NEMA 1 •
Electronic
Compartment:
IP20
3 Lbs (1.4 kg)
12-months from date of shipment
Revision Level: C
ECO 8029
The solid state electrolyte depletes with high oxygen
File Name: MN-A-0008
Appendix B - Specifications
Page 5-3