4.1.2.11.6 Force minimum Current Output to O
This menu sets the Oxygen concentration at which the Analog Current output is at minimum range (section
4.1.2.11.2).
This setting is entered in percent oxygen increments and can be anywhere from
0.0 % to 100.0 %. This menu is accessed from the Analog Output Setup menu by typing "6" on the RS-232
terminal. . Contact factory before changing these settings.
4.1.2.11.7 Force Maximum Current Output to O
This menu sets the Oxygen concentration at which the Analog Current output is at maximum range (section
4.1.2.11.2).
This setting is entered in percent oxygen increments and can be anywhere from
0.0 % to 100.0 %. This menu is accessed from the Analog Output Setup menu by typing "7" on the RS-232
terminal. . Contact factory before changing these settings.
4.1.2.12
Display/Auto-Range Setup
The RS-232 Display/Auto-Range Setup menu U14 (see Figure 18)provides access for the user to map the
display and Analog output range scale(s) of the Model 5100E to suit the application (Appendix E – Range /
Analog output Chart). The Analog Output Range may be set to 1 (fixed range 0-1 %) • 2 (fixed range 0-10
%) • 3 (fixed range 0-25 %) • 4 (fixed range 0-50 %) • 5 (fixed range 0-100 %) • 8 (low auto-range) • 9 (high
auto-range) • 10 (full auto-range). This menu is accessed from the Setup Main Menu by typing "4" or "G"
on the RS-232 terminal. To navigate backwards, use the <Esc> or "Q" key on the RS-232 terminal. . Refer to
the Software Setup Log (that is supplied by the factory) for the configuration details of your Model 5100E
Figure 18 – Display/Auto Range Setup Menu
Manual Part Number:
C5-06-4900-08-0
Concentration
2
Concentration
2
Revision Level: C
ECO 8029
File Name: MN-A-0008
System Setup
Page 4-13