Light Mode Service Menu; Flow Sensor Service Menu - LaboGene Mars Service Manual

Class 2-biohazard safety cabinet or cytotoxic cabinet with automatic control system microprocessor based
Table of Contents

Advertisement

Edition 3 April 2012
Once the correct mode is selected press ENTER● to accept the selection and this will return the display to the Service
Mode.
7.3.5.7 LIGHT MODE Service sub menu
In the light mode menu, 3 options can be selected for the fluorescent light settings during REDUCED
Once the correct mode has been set, press the ENTER ● button and this will return to the display to the main menu.
7.3.5.8 FLOW SENSOR service sub menu
The FLOW SENSOR menu allows the setting of the type of flow sensor and there are 3 sub menus which can be
selected.
SAFE mode means the Fans cannot be turned off unless power to the cabinet is
turned off. They will run continuously but REDUCED SPEED can be selected.
FULL mode means the fans will run only at full speed and there is no reduced
speed option.
Press ENTER to enter the LIGHT MODE and the sub menus are activated.
The SAFE mode should be selected when no fluorescent light is to be on in the
work area when the REDUCED SPEED fan is running in the cabinet or there is an
alarm. Use the ▼ and ▲ buttons to scroll to next mode if this is not required.
The AUTO mode should be selected if lights are to run continuously.
The NORMAL mode should be selected if the lights are to remain ON during
REDUCED SPEED and if there is an alarm
Press ENTER ● to enter and activate the sub menu options.
Page 33/59