LaboGene Mars Service Manual page 13

Class 2-biohazard safety cabinet or cytotoxic cabinet with automatic control system microprocessor based
Table of Contents

Advertisement

Edition 3 April 2012
Mars/Mars Pro Service Manual
10. PIR sensor
. 9.001.000.087
The PIR is used when fully automatic operation of the cabinet is required.
The cabinet is run at reduced speed and when the operator starts to work in the cabinet the
PIR sensor detects the movement and automatically sets the fans to run at normal speed. If no movement is
detected for 10 minutes then the fans automatically revert to reduced speed