Maintenance by the operator
Monthly operator maintenance work
Switch off the instrument completely.
Further operator maintenance work
Backup of data (complete registry =
program and data).
Energy channel recalibration (MCA),
depending on the results of the control
sample measurement.
Global recalibration, depending on the
results of the control sample measurement.
(Should the MCA recalibration procedure
fail to bring improvement.)
Clean the sample tray.
Replace the "liquid protect cups" protective
film.
Clean the sample rotation drive (do not
lubricate!)
— 30
This is a procedure requested by the
software program every 30 days. Switch off
the instrument and leave it off for 1 minute
before switching it on again. Safety devices
are automatically checked for correct
functioning whenever the instrument is
restarted.
Then, carry out an energy channel
recalibration procedure (MCA) (see Chapter
6.2.1).
Directory C:\Spectro Smart Studio\XRF
Analyzer Pro
Place the recalibration aid at the
measurement position and start the energy
channel recalibration procedure (MCA)
(Chapter 6.2.1).
Place the recalibration aid in the instrument
and start the global recalibration procedure
(Chapter 6.2.2).
See chapter 6.3
See Chapter 6.5.
Use a soft brush.