3
Installation
Install KingFisher Presto
Install the Magnet Head
To insert or change the magnet head:
1. Find the Change magnet head protocol by pressing the Select button on the instrument holder.
2. Press the Start button.
The magnet head holder rises.
3. Insert the magnet head:
a. Lift the magnet head carefully from its storage box by holding it from the top.
b. Check that the three pins on the top of the magnet head go to the slots of the magnet head
c. Push the magnet head slowly until the pins clicks into their notches.
4. Press Start to confirm the end of the protocol.
Install the Heating Block
To insert or change the heating block:
1. Find the Change heating block protocol by pressing the Select button.
2. Press Start.
The tip comb holder rises.
3. Insert the heating block:
a. Set the heating block into its slot.
20
Thermo Scientific KingFisher Presto User Manual
holder.
Figure 12. Inserting the magnet head
Thermo Fisher Scientific