Troubleshooting and diagnostics
January 2020
7. The example screen shows the Fit6/Path2 status is XShift Error . The XShift
8. Examine the data shown on the corresponding Faults screen to determine if a
9. Contact your Emerson Customer Care Representative
7.2.2
Diagnostics menu screen
The Diagnostics screen lists those main components of the analyzer where problems can
be diagnosed using the BIST system. The Diagnostics screen also enables you to check on
the status and, where appropriate, the values of any of the listed components.
The Diagnostics menu screen can be accessed from the Main Menu screen.
This gives the operator access to a number of options when checking the overall health of
the analyzer.
Procedure
1. Where more options are available than can be displayed on the screen use the
2. Pressing the Forward key
70
value is not within the defined tolerance. When rectified, the status will change to
.
OK
Figure 7-4: Fault Status Details Screen
component has failed or if it is operating outside of its correct parameters.
(cascade.support@emerson.com) if you need further assistance with a
component's Fault Status.
or
keys to scroll up or down onto the next page. The screen will continually wrap
around until the option required is highlighted.
in the Diagnostics/Config example below. Pressing the screen Refresh key at any
time will refresh the current display screen if required by the operator.
CONFIG
gives the status of the analyzer CONFIGURATION file.
SENSOR
gives the status of the analyzer SENSOR(s) engine.
will move to the next screen where available as shown
Reference Manual
00809-0100-4244
Rosemount CT4400