Emerson Rosemount CT4400 Reference Manual page 28

Continuous gas analyzer
Table of Contents

Advertisement

Install
January 2020
Table 2-2: Analog Outputs Full Rack System, 25-Way Socket
Pins
1 & 14
2 & 15
3 & 16
4 & 17
5 & 18
6 &.19
7 & 20
8 & 21
9 & 22
10 & 23
11 & 24
12 & 25
28
Figure 2-6: Analog Connector 25-Way Socket
Function
Analog channel 0, set in config file (4-20 mA, SELV)
Analog channel 1, set in config file (4-20 mA, SELV)
Analog channel 2, set in config file (4-20 mA, SELV)
Analog channel 3, set in config file (4-20 mA, SELV)
Analog channel 4, set in config file (4-20 mA, SELV)
Analog channel 5, set in config file (4-20 mA, SELV)
Analog channel 6, set in config file (4-20 mA, SELV)
Analog channel 7, set in config file (dry contacts: 1A, 30 Vdc)
Status output 1 (check function), optional (dry contacts: 1A, 30
Vdc)
Status output 2 (maintenance required), optional (dry contacts:
1A, 30 Vdc)
Status output 3 (out of specification), optional (dry contacts: 1A,
30 Vdc)
Status output 4 (failed), optional (dry contacts: 1A, 30 Vdc)
Reference Manual
00809-0100-4244
Measurement
(position on display)
Sample gas
measurement 1
Sample gas
measurement 2
Sample gas
measurement 3
Sample gas
measurement 4
Sample gas
measurement 5
Sample gas
measurement 6
Sample gas
measurement 7
Sample gas
measurement 8
N/A
N/A
N/A
N/A
Rosemount CT4400