Calibration - Emerson Rosemount CT4400 Reference Manual

Continuous gas analyzer
Table of Contents

Advertisement

Gas calibration
January 2020
CALIBR PARAMS
VERIFY PARAMS
STREAM SWITCH
Pressing the Back key
any time will return to the Main Menu screen.
6.3

Calibration

Calibration flows the known cylinder gas through the analyzer and then adjusts the
readout until the measurement matches the cylinder. The analyzer then applies this
adjustment to all measurements until the next calibration. There are two circumstances in
which you may want to calibrate:
1. After you verify gas concentrations (see
2. At regular intervals, such as once a month.
6.3.1
Zero calibration
The zero calibration procedure adjusts for any zero offset in the measurement. Zero
calibration is done by calibrating the analyzer measurements against a known sample gas
using the following procedure.
Prerequisites
Run the analyzer at a stable temperature for at least thirty minutes prior to commencing
this procedure.
Procedure
1. Use nitrogen gas of "instrument gas" purity as the zero calibration gas.
2. Ensure that a pressure regulator is connected to the nitrogen gas bottle.
3. Connect a hose from the nitrogen gas bottle through a T-piece to the sample supply
4. Connect an excess flow line to the unused port on the T-piece and route the excess
5. On the display controller of the analyzer, browse to the Main menu as described in
50
allows configuration of the purge time for Gas Calibration.
allows configuration of the purge time for Gas Verification.
not an option available for the Rosemount CT4400 analyzer.
will return to the previous screen. Pressing the Home key
the analyzer readings are outside of tolerance.
port on the rear of the analyzer.
flow line to a suitable extractor.
Gas service menu
definitions.
Verifying gas
concentrations) and find that
Reference Manual
00809-0100-4244
at
Rosemount CT4400