Performance Tests
Stop shows test complete
Falling edge at center crossing
1–80
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
e. Press the side-menu V-Bars.
f.
Use the general purpose knob to move the left cursor to the right until
the @ measurement reads -32 ns.
g. Press SELECT.
h. Use the General Purpose knob to adjust the right cursor until the delta
measurement reads 60 ns.
i.
CONFIRM that the width of the falling edge of the trace (at the center
crossings) falls between the cursors. See Figure 1–27.
j.
Press DISPLAY.
k. Press the main-menu Format/RO.
l.
Press the side–menu Display 'T' @ Trigger Point until you select ON.
Figure 1–27: Jitter test when completed – TDS 754D shown
5. Check Triggered Signal Range.
Set up oscilloscope for Triggered Signal Test.
a. Press MORE.
b. Press WAVEFORM OFF.