Figure 1-3: Universal Test Hookup For Functional Tests - Tds 600C - Tektronix TDS 500D Technical Reference

Digitizing oscilloscopes performance verification and specifications
Hide thumbs Also See for TDS 500D:
Table of Contents

Advertisement

Brief Procedures
Verify All Input Channels
1–8
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Equipment
One probe such as the P6243, P6245, P6249, P6139A
required
Prerequisites
None
1
P6139A probe is not appropriate for the TDS694C/794D oscilloscopes.
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground (typically black) to PROBE COMPENSA-
TION GND. If using a P6243, P6245 or P6249 probe, you may want to
attach a Y-lead connector and two SMD KlipChips as shown in
Figure 1–3.
Figure 1–3: Universal test hookup for functional tests – TDS 600C shown
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the main-menu button Recall Factory Setup.
H Press the side-menu button OK Confirm Factory Init.
1
, or P6339A
Signal
Gnd

Hide quick links:

Advertisement

Table of Contents
loading