Table 2-17: Typical Characteristics - Time Base System - Tektronix TDS 500D Technical Reference

Digitizing oscilloscopes performance verification and specifications
Hide thumbs Also See for TDS 500D:
Table of Contents

Advertisement

Typical Characteristics
Table 2–16: Typical characteristics — Signal acquisition system (cont.)
Name
Step Response Settling Errors
Step Response Settling Errors
TDS 694C/794D
1
The limits given are for the ambient temperature range of 0_C to +30_C, TDS 694C 4_C to +30_C. Reduce the upper
bandwidth frequencies by 5 MHz for the TDS 600C or by 2.5 MHz for the TDS 500D/700D/714L for each _C above +30_C. For
the TDS 694C and TDS 794D reduce the upper bandwidth frequencies by 10 MHz for each _C above +30_C.
2
The numbers given are valid 0_C to +30_C and will increase as the temperature increases due to the degradation in
bandwidth. Rise time is calculated from the bandwidth. It is defined by the following formula:
TDS 600C Rise Time (ns) +
Note that if you measure rise time, you must take into account the rise time of the test equipment (signal source, etc.) that
you use to provide the test signal. That is, the measured rise time (RT
the rise time of the test signal source (RTgen) according to the following formula:
+ RT
) RT
2
2
RT
m
i
3
The values given are the maximum absolute difference between the value at the end of a specified time interval after the
midlevel crossing of the step and the value one second after the midlevel crossing of the step, expressed as a percentage
of the step amplitude.
Table 2–17: Typical characteristics — Time base system
Name
Accuracy, Delta Time Measurement
(TDS 500D/700D Option 1G and
TDS 714L)
Conditions for accuracy listed at right
are: Single Shot, Sample or HiRes
mode with Full Bandwidth selected.
2–32
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description
Volts/Div setting
1 mV/div – 100 mV/div
101 mV/div – 1 V/div
1.01 V/div – 10 V/div
10 mV/div – 100 mV/div
101 mV/div – 1 V/div
450
BW (MHz)
2
gen
Description
The limits are given in the following table for signals having amplitude greater than
5 divisions, reference level = 50%, filter set to (sinX/X), acquired at 5 mV/div or greater.
For the TDS 500D/700D Option 1G and the TDS 714L, pulse duration < 10 div. Channel
skew not included.
For the Single Shot condition, 1.4 ≤ T
S is the sample rate and T
Extra error in the measurement will occur for two-channel measurements due to
channel-to-channel skew. This is described elsewhere in these specifications.
Time measurement accuracy
±((0.15 B sample rate) + (25 ppm × | Reading |))
± Step amplitude
≤2 V
≤20 V
≤200 V
≤1.5 V
≤3 V
TDS 500D 694C 700D Rise Time (ns) +
) is determined by the instrument rise time (RT
m
B S ≤ 4, where
r
is the displayed rise time.
r
3
Settling error (%)
at
20 ns
100 ns 20 ms
≤0.5% ≤0.2% ≤0.1%
≤1.0% ≤0.5% ≤0.2%
≤1.0% ≤0.5% ≤0.2%
≤0.5% ≤0.2% ≤0.1%
≤1.0% ≤0.5% ≤0.2%
400
BW (MHz)
) and
i

Hide quick links: