Limits; Table 2-13: Warranted Characteristics - Output Ports, Probe - Tektronix TDS 500D Technical Reference

Digitizing oscilloscopes performance verification and specifications
Hide thumbs Also See for TDS 500D:
Table of Contents

Advertisement

Warranted Characteristics
Table 2–12: Warranted characteristics — Triggering system (Cont.)
Name
Accuracy (Time) for Pulse-Glitch or
Pulse-Width Triggering
Input Signal Sync Amplitude for Stable
Triggering, NTSC and PAL modes
(Option 05 Video Trigger)
Jitter (Option 05 Video Trigger)
1
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on
the selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not
"roll" across the screen on successive acquisitions. The TRIG'D LED stays constantly lighted when the SEC/DIV setting
is 2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
2
For Slew Rate Triggering, this is the minimum transition time, defined to be the time the user's signal spends between
the two trigger threshold settings.
Table 2–13: Warranted characteristics — Output ports, probe compensator, and power requirements
Name
Logic Levels, Main- and Delayed-Trigger
Outputs
Output Voltage and Frequency,
Probe Compensator
Output Voltage, Signal Out (CH 3
Not on TDS 694C
Source Voltage
Source Frequency
Power Consumption
TDS 694C
1
CH 3 signal out is present at the rear panel if CH 3 (AUX 1 on the TDS 680C) is selected as the trigger source for the main
and/or delayed trigger systems. It is not available when a channel other than CH3 (AUX 1 on the TDS 680C) is the source
for the Video Trigger when Option 05 is installed.
2–24
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description
Time range
Accuracy
±(20% of setting + 0.5 ns)
2 ns to 500 ns
±(100 ns + 0.01% of Setting)
520 ns to 1 s
Field selection "Odd", "Even", or "All": 0.6 division to 4 divisions
Field selection "Numeric": 1 division to 4 divisions (NTSC mode)
60 ns
on NTSC or PAL signal
p-p
Description
Characteristic
Vout (HI)
Vout (LO)
Characteristic
Output Voltage
Frequency
For TDS 600C: 20 mV/division ±20% into a 1 MW load;
1
)
10 mV/division ±20% into a 50 W load
For TDS 500D/700D/714L:
22 mV/division ±20% into a 1 MW load;
11 mV/division ±20% into a 50 W load
100 to 240 VAC
, continuous range, CAT II
RMS
45 Hz to 440 Hz
≤350 W (450 VA)
≤450 W (500 VA)

Limits

≥2.5 V open circuit; ≥1.0 V into a 50 W
load to ground
≤0.7 V into a load of ≤4 mA;
≤0.25 V into a 50 W load to ground
Limits
0.5 V (base-top) ±1% into a ≥50 W load
1 kHz ±5%