Table 1-3: Dc Voltage Measurement Accuracy - Tektronix TDS 500D Technical Reference

Digitizing oscilloscopes performance verification and specifications
Hide thumbs Also See for TDS 500D:
Table of Contents

Advertisement

Performance Tests
1–40
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 1–3: DC Voltage measurement accuracy
Position
Scale
Offset
setting
setting
setting
(Divs)
5 mV
–5
+1 V
+5
–1 V
200 mV –5
+10 V
+5
–10 V
1.01 V
–5
+10 V
+5
–10 V
10 mV
–5
+0.5 V
+5
–0.5 V
200 mV 0
+5 V
0
–5 V
1 V
0
+2.5 V
0
–2.5 V
d. Display the test signal:
H Press VERTICAL MENU. Press the main-menu button Position.
H Use the keypad to set vertical position to –5 divisions (press –5, then
ENTER, on the keypad). The baseline level will move off screen.
H Press the main-menu button Offset.
H Use the keypad to set vertical offset to the positive-polarity setting
listed in the table for the current vertical scale setting. The baseline
level will remain off screen.
H Set the generator to the level and polarity indicated in the table for
the vertical scale, position, and offset settings you have made. The
DC test level should appear on screen. (If it doesn't return, the DC
accuracy check has failed for the current vertical scale setting of the
current channel).
e. Measure the test signal: Press CLEAR MENU. Read the measurement
results at the Mean measurement readout. See Figure 1–6.
TDS 600C
Generator
(except TDS694C)
setting
Accuracy limits
+1.040 V
+1.0329 V to +1.0471 V
–1.040 V
–1.0471 V to –1.0329 V
+11.6 V
+11.4165 V to +11.7835 V
–11.6 V
–11.7835 V to –11.4165 V
+18V
+17.102 V to +18.899 V
–18 V
–18.899 V to –17.102 V
TDS694C
Accuracy limits
+0.58 V
+575.5 mV to +584.5 mV
–0.58 V
–584.5 mV to –575.5 mV
+5.0 V
+4.828 V to +5.172 V
–5.0 V
–5.172 V to –4.828 V
+5.0 V
+4.738 V to +5.263 V
–5.0 V
–5.263 V to –4.738 V
TDS 500D/700D/714L
(except TDS794D)
Accuracy limits
+1.0355 V to +1.0445 V
–1.0445 V to –1.0355 V
+11.5195 V to +11.6805 V
–11.6805 V to –11.5195 V
+17.621 V to +18.379 V
–18.379 V to –17.621 V
TDS794D
Accuracy limits
+575.5 mV to +584.5 mV
–584.5 mV to –575.5 mV
+4.9405 V to +5.0595 V
–5.0595 V to –4.9405 V
+4.794 V to +5.206 V
–5.206 V to –4.794 V

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents