Option 05 Video Trigger Checks; (Not Available On Tds694C Or Tds794D) - Tektronix TDS 500D Technical Reference

Digitizing oscilloscopes performance verification and specifications
Hide thumbs Also See for TDS 500D:
Table of Contents

Advertisement

Option 05 Video Trigger Checks

(Not Available on TDS694C or TDS794D)

Check Video Trigger
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Equipment
PAL signal source (Item 21)
required
60 Hz sine wave generator (Item 22)
Pulse generator (Item 23)
Two 75 W cables (Item 24)
Two 75 W terminators (Item 25)
One BNC T connector (Item 7)
One precision 50 W coaxial cable (Item 5)
50 W terminator (Item 4)
Prerequisites
See page 1–17. These prerequisites include running the signal path
compensation routine.
1. Set up the oscilloscope to factory defaults by completing the following steps:
a. Press save/recall SETUP.
b. Press the main-menu Recall Factory Setup.
c. Press the side-menu OK Confirm Factory Init.
d. Wait for the Clock Icon to leave the screen.
e. CONFIRM the oscilloscope is setup as shown below.
Channel: CH1
Volt/div: 100 mV
Horizontal scale: 500 ms/div
2. Set up the oscilloscope for TV triggers by completing the following steps:
a. Press TRIGGER MENU.
b. Press the main-menu Type pop-up until you select Video.
c. Press the main-menu Standard pop-up until you select 625/PAL.
d. Press the main-menu Line.
e. Use the keypad to set the line number to 7 (press 7, then ENTER).
f.
Press VERTICAL MENU.
g. Press the main-menu Bandwidth.
Performance Tests