Check Character Generator; Check Polarity Indication; Check Dc Voltage Measurement Accuracy; Check Dc Current Measurement Accuracy - Tektronix 213 DMM Instruction Manual

Oscilloscope with options
Table of Contents

Advertisement

Calibration-213 Service
g. Set Standard to +3.000 volts.
h. CHECK-Reading displays an overrange condition
(scrambled character segments or B's).
4. CHECK CHARACTER GENERATOR
a. Set the DC Voltage Standard to +1.005 volts.
b. Rotate the Standards 100 mV selector from
O
through
9 ( 1.005 to 1.905).
c. CHECK-Millivolt digit on the readout displays a 0
through 9 (+1.005 to +1.905).
5. CHECK POLARITY INDICATION
a. Note the polarity of the reading on the display.
b. Reverse the banana plug connector at the Standard.
c. CHECK-The display reverses polarity sign.
d. Reset banana plug connector for positive inputs.
6. CHECK DC VOLTAGE MEASUREMENT ACCURACY
a. Set VOL TS-mA/DIV and the DC Voltage Standard as
shown in Table 6-6.
b. CHECK-Reading is within limits shown in Table 6-6.
c. Disconnect the Standard.
7. CHECK DC CURRENT MEASUREMENT ACCURACY
a. Set: VOLTS-mA/DIV
mA
0.1 (blue scale)
Button in
b. Connect the DC Current Standard to the mA-n and
common jacks using two banana-plug to BNC adapters
and a cable.
6-14
TABLE 6-6
DC Voltage Measurement Accuracy
VOL TS-mA/DIV
DC Voltage
Setting
Standard
Reading
Limits
(blue scale)
Setting
0.1
+0.1000 V
+. 1000
+.0996 to +.1004
-0.1000 V
-.1000
-.0996 to -.1004
10
+20.00 V
+20.00
+19.96 to +20.04
-20.00 V
-20.00
-19.96 to -20.04
100
+200.0 V
+200.0
+199.6 to +200.4
-200.0 V
-200.0
-199.6 to -200.4
1000
+750 V
+0750
+0747 to +0753
-750V
-750
-0747 to -0753
c. Set VOL TS-mA/DIV and the Standard as shown in
Table 6-7.
d. CHECK-Reading is within limits shown in Table 6-7.
e. Disconnect the Standard.
TABLE 6-7
DC Current Measurement Accuracy
VOL TS-mA/DIV
DC Current
Setting
Standard
Reading
Limits
(blue scale)
Setting
0.1
all zeros
.0000
-.0003 to +.0003
0.1
+0.1 mA
+. 1000
+.0992 to +.1008
1
+1mA
+1.000
+0.994 to +1.006
10
+10mA
+10.00
+09.94 to +10.06
100
+100 mA
+100.0
+099.4 to +100.6
1000
+1 A
+1'000
+0994 to + 1006
8. CHECK/ADJUST OHMS CALIBRATION
a. Set: n
VOLTS-mA/DIV
Button in
10 kr2
@

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents