Assemble the Stand

Thermo Scientific

Start-Up

Section 3
Caution Correct assembly and installation is essential for proper start-up
s
and operation.
For a unit without a stand, place the cabinet onto a stable structure
capable of supporting two times the unit weight. Make sure the weight is
not resting on the drain valve or nipple.
For a unit with a stand, first assemble the stand per the instructions
included with the stand. Then fit the cabinet onto the stand. See steps
below.
Caution Make sure each leveler is turned all the way in, to avoid bending
an edge during installation.
1. Locate the four telescoping legs in the side panel packaging. Install
adjustable leveler from the hardware bag into the swaged end of the
telescoping legs (if not already installed).
NOTE: Make sure leveler is screwed all the way in, before installation.
Using a hammer, tap the leveler into place.
2. Locate the left and right side panels. Install the telescoping leg
assemblies into the side panels as shown in Figure 3-1.
3. The height-adjustable stand can be set to a height range of
26.8 - 34.7 in. (680 - 880 mm ). To set the stand to the desired
height, pull the legs downward out of the channels to the desired
height. Lock the height adjustment in place by installing two
lockwashers/screws per channel to the leg.
Note All lockwashers are a wedgelock type. The wedges fit together with
the smoother sides toward the unit and the screw. See Detail.
4. Locate the two crossmembers from the packaging. Note they are
packaged separately from the side panels. With the label on the
crossmember to the outside (Figure 3-1), install the crossmembers onto
the retaining tabs of the side panels. Place lockwashers on screws, then
install all screws by hand, before tightening to secure together.
5. To install the cabinet frame to the stand, insert four flatwashers and
Allen screws loosely into the threaded holes at the bottom of the
cabinet.