Section 8
Certification Testing
Testing Information
(continued)
8-4
1300 Series A2
Inflow Area:
• 3ft Models (10" window opening work position) 1320, 1321, 1322,
1323, 1330, 1331, 1332, 1333: Front opening 2.46 sq ft
3ft Models (8" window opening work position) 1339, 1340, 1341,
1342, 1390, 1391, 1392, 1393: Inflow area for 8" models: 1.97 sq ft
• 5ft Models (10" window opening work position) 1368, 1369, 1370,
1371, 1380, 1381, 1382, 1383: Front opening 4.10 sq ft
5ft Models (8" window opening work position) 1350, 1351, 1352,
1353, 1366, 1367, 1372, 1373: Inflow area for 8" models: 3.28 sq ft
• 4ft Models (10" window opening work position) 1345, 1346, 1354,
1355, 1356, 1375, 1376, 1384, 1385, 1386: Front opening
3.28 sq ft
4ft Models (8" window opening work position) 1305, 1306, 1325,
1326, 1335, 1336, 1395, 1396: Inflow area for 8" models: 2.62 sq ft
• 6ft Models (10" window opening work position) 1347, 1348, 1357,
1358, 1359, 1377, 1378, 1387, 1388, 1389: Front opening
4.92 sq ft
6ft Models (8" window opening work position) 1307, 1308, 1327,
1328, 1337, 1338, 1397, 1398: Inflow are for 8" models: 3.93 sq ft
Calculation:
• Inflow Velocity (V1) = Inflow Air Volume (V2)/Actual Inflow Area (A2)
Acceptance:
• 100 – 110 FPM
Correction measures:
• Enter the service level and activate S2, change S2 until the desired
velocity is reached and save the new value.
• If the velocity can't be adjusted due to clogging of the filters, change the
exhaust filter.
• If the filter is in order, change the exhaust blower.
Note Check of alarm limits: If S1 or S2 are changed, the alarm limits S3
and S4 must be checked and saved again.