LT1317X1 • 26/04/10
Super-Nuova
Hot plates
operaTIoN MaNuaL
aND parTS LIST
SerIeS 1317 and 1337
Model #
Hp131720-33
Hp131725
Hp131724
Hp133730-33
Hp133735
Hp133734
Voltage
Description
220-240
120
100
220-240
120
100
7x7
7x7
7x7
10x10
10x10
10x10

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Super-Nuova 1317 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Thermo Scientific Super-Nuova 1317 Series

  • Page 1 Super-Nuova Hot plates operaTIoN MaNuaL aND parTS LIST SerIeS 1317 and 1337 Model # Voltage Description Hp131720-33 220-240 Hp131725 Hp131724 Hp133730-33 220-240 10x10 Hp133735 10x10 Hp133734 10x10 LT1317X1 • 26/04/10...
  • Page 2: Table Of Contents

    Table of Contents Safety Information ..............................3 alert Boxes ..............................3 Warnings ..............................3 General Specifications ............................5 Heating Specifications ..........................6 environmental Conditions ..........................7 Declaration of Conformity ..........................7 Introduction................................8 General usage..............................8 unpacking and Installation ..........................9 unpacking ..............................9 Installation ..............................9 principles of operation ............................11 operation ................................12 power Button ..............................12 Setting the Temperature ..........................12 Controlling Solution Temperature with an external probe................13...
  • Page 3: Safety Information

    Safety Information Your Thermo Scientific Super-Nuova Hot plate has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. For safe operation, please pay alert Signals attention to the alert signals throughout the manual. Warning: These products should be used only under the Warning operating conditions specified in the operating Manual.
  • Page 4 afeTy NforMaTIoN Keep top surface clean. use a non-abrasive cleaner. alkali spills, hydrofluoric acid spills or phosphoric acid spills may damage top and lead to thermal failure. unplug unit and remove spills promptly. Do not immerse unit for cleaning. replace the top immediately if damaged by etching, scratching or chipping.
  • Page 5: General Specifications

    General Specifications 7" X 7" Model Number HP131725 HP131720-33 HP131724 overall Dimensions in. (cm) Width 8.2” (20.8 cm) 8.2” (20.8 cm) 8.2” (20.8 cm) Height 3.8” (9.7 cm) 3.8” (9.7 cm) 3.8” (9.7 cm) Depth 13.0” (33.0 cm) 13.0” (33.0 cm) 13.0”...
  • Page 6: Heating Specifications

    eNeral PeCIfICaTIoNS Heating Specifications Top Plate Surface - Solid Ceramic Temperature range: 1°C - 370°C (34°F - 698°F)* * This hot plate does not cool. The minimum temperature is 1°C if used in a cold room below 1°C. 7” x 7” 10”...
  • Page 7: Environmental Conditions

    eNeral PeCIfICaTIoNS environmental Conditions operating: 0°C to 27°C; 20% to 80% relative humidity, non-condensing. Installation category II (overvoltage) in accordance with IeC 664. pollution degree 2 in accordance with IeC 664. altitude Limit: 2,000 meters. Storage: -25°C to 65°C 10% to 85% relative humidity Declaration of Conformity (for 220-240 volt, -33 Ce models only) We hereby declare under our sole responsibility that this product conforms with the technical requirements of...
  • Page 8: Introduction

    Introduction please read all the information in this manual before operat- ing the unit. Your Super-Nuova hot plate is a heating plate designed for laboratory procedures requiring precise control of tem- perature. each Super-Nuova model includes a digital dis- play for monitoring temperature. The hot plate is capable of producing accurately controlled top plate temperatures from 1°C through 370°C.
  • Page 9: Unpacking And Installation

    unpacking and Installation unpacking Warning remove your Super-Nuova hot plate from the carton. use a properly grounded electrical Inspect to ensure that the unit has not been damaged outlet of correct voltage and current during shipment. If the unit appears to have sustained handling capacity.
  • Page 10: Installation

    Installation Set the unit on a flat stable surface at least 12" away from combustible materials, and plug the cordset into a proper- ly grounded electrical outlet of correct voltage and current handling capacity.
  • Page 11 NPaCkING aND NSTallaTIoN RS232 Port Connections Cord retention clip Remote Probe Connection Temperature CAUTION! HOT TOP Temperature Display Indicator Set Button HEAT HEAT POWER POWER Power Button Writable Preset CAUTION! UTION! Timer Surface* HOT TOP HOT TOP Indicator Lights EXTERNAL EXTERNAL PROBE PROBE...
  • Page 12: Principles Of Operation

    principles of operation each Super-Nuova unit utilizes the latest microproces- sor technology to deliver the most reliable, accurately controlled, ceramic top hot plate on the market. Your Super-Nuova hot plate has an electronic, closed- loop feedback control which will accurately maintain tem- perature setpoints from 1°C through 370°C.
  • Page 13: Operation

    operation Warning use caution when heating volatile power Button materials; top surface and element The Super-Nuova unit has a power button located on can reach the “Flash point the front control panel. The unit may be powered on by Temperature” of many chemicals. pressing the power button or aNY button on the control These hot plates are not explosion panel.
  • Page 15 PeraTIoN Controlling Solution Temperature using external Note probe The top plate temperature will be higher than your solution tempera- To control the solution temperature plug the included ture. probe or an ungrounded Type K thermocouple probe into the probe receptacle located on the right side of the unit. Be careful to observe the correct polarity of the probe connector when inserting probe.
  • Page 16: Using Preset Programs

    PeraTIoN using preset programs The Super-Nuova hot plate features the option of pre- setting 4 setpoints into its memory. To program a preset, adjust the heat features to your desired setpoints. press and hold one of the presets and the unit will beep indicat- ing those setpoints have been stored in memory.
  • Page 17: Setting The Timer

    PeraTIoN Setting the Timer The timer function on the Super-Nuova can be set to shut off heating. The timer can be set for a desired number of hours (1, 2, 4, 8 or 12) or for a non-specific period of time (Infinity symbol).
  • Page 18: Rs232 Output

    PeraTIoN rS232 output an rS232 port is available on the Super-Nuova to output elapsed time and temperature data. The data is printed once every second when the unit is powered on (displays on). a comma separates each piece of data so that it may easily be imported into a spreadsheet as CSv (Comma Separated values).
  • Page 19: Using Hyperterminal W/Rs232

    PeraTIoN using HyperTerminal w/rS232 With HyperTerminal you can store the rS232 output data of a Super-Nuova on a pC. NoTe: other terminal programs that capture rS232 data can be used as well. In order to follow this procedure, you must have: •...
  • Page 20: To Capture Data W/Hyperterminal

    PeraTIoN You will be able to run HyperTerminal with this setup in the future by just double-clicking the icon if you put it on the Desktop. The setup of HyperTerminal is now complete, you can view the past data, but the program will only allow you to see the past 500 lines in its buffer.
  • Page 22: Service And Calibration

    Service and Calibration Service Menu The Service Menu has many features that will allow a user to customize their unit. a table of the features available in the Service Menu is given below. The features are given in order of their appearance in the Service Menu when rotating the center knob in a clockwise rotation.
  • Page 23: Over-Temperature Set Point - Osp

    erVICe aND alIbraTIoN over-temperature Set point - Note For most routines, the poWer key This option is only available for units that have heating can be used to return to a previous functionality. The over-temperature Set point is the set menu if desired.
  • Page 24: Temperature Probe Calibration - Cal

    erVICe aND alIbraTIoN Temperature probe Calibration - CaL This section applies only to the calibration of the temperature probe system. The hot plate surface cannot be calibrated. This method is only applicable to units that have heating func- tionality. There are two types of calibration available. The choice of which calibration to choose will depend on the equipment that is available.
  • Page 25: Oil Bath Method Of Calibration

    erVICe aND alIbraTIoN oil Bath Method of Calibration With the probe connected to the unit, insert the probe connected to the unit to be calibrated into a bath that has stabilized at the desired calibra- tion temperature. allow sufficient time for the probe to stabilize.
  • Page 26: System Method Of Calibration

    erVICe aND alIbraTIoN System Method of Calibration Select an independent temperature probe to be placed in the load, along with the unit tempera- ture probe before continuing with calibration. Select the load to be calibrated, and place on top of the hot plate. Make sure to have probe plugged into the unit and placed in the load.
  • Page 27: Thermometer Mode - Prb

    erVICe aND alIbraTIoN and the display(s) will show the stirring control set point. The speed may be adjusted by using the center knob. The adjustable range is 50 – 999, and also oFF (zero). When the adjustment is completed, press the SeT key under the STIr display to approve the selection.
  • Page 28: Timer Shutdown - End

    erVICe aND alIbraTIoN remove your finger from the poWer key. The display will change to oSP. rotate the center knob until the display reads prb. press the SeT key under the HeaT display to accept, and the display will change to oFF. rotate the center knob one position to change the display to on and press the SeT key under the HeaT display again.
  • Page 29: Probe Temperature Limit - Pl

    erVICe aND alIbraTIoN the display to the method of choice and press the SeT key under either display again. The unit will return to the previous menu. Select another feature to change, or press the poWer key again to return to the off mode. probe Temperature Limit - pL This feature is available for units with heating functionality and is only applicable when using a probe.
  • Page 30: Probe Response - Pr

    erVICe aND alIbraTIoN probe response - pr This feature is available for units with heating functionality and is only applicable when using a probe. The purpose of this feature is to select the minimum time required for a temperature change to be detected before signaling a probe out of Solution error (e03).
  • Page 31: Error Disable - Err

    erVICe aND alIbraTIoN Therefore the oSP setting would need to be at least 300°C. The probe circuit is also affected by the selection of oSp. Selection of Ul would allow the set point to be unlimited and would not be affected by oSP setting. To modify/view the Set point Limit follow the instructions given below: plug the unit into the appropriate power, but do...
  • Page 32: Factory Defaults - Def

    erVICe aND alIbraTIoN a single beep will be heard and you may remove your finger from the poWer key. The display will change to oSP. rotate the center knob until the display reads err. press the SeT key under the display to accept, and the display will change to Clr (re- enable all disabled errors).
  • Page 33: Model Selection - Sel

    erVICe aND alIbraTIoN to change the display to yeS and press the SeT key. The unit has now been returned to the factory default settings. The unit will return to the previ- ous menu. Select another feature to change, or press the poWer key again to return to the off mode.
  • Page 34: Flash Upgrade - Fls

    erVICe aND alIbraTIoN Flash upgrade - FLS This feature is available for all units and is designed to allow a firmware upgrade via rS232 port to be applied without exchanging controls. NoTe: Special software and an rS232 cable are required to complete this proce- dure.
  • Page 35: Error Codes

    Troubleshooting error Codes errors e01 through e07 are heating errors. error Handler will lock out heating functions if heating error is detected. Stirring functionality is unaffected. If the condition that caused the error is no longer present, pressing the poWer button or unplugging the unit will clear errors e01-e07. y t i .
  • Page 36 roUbleSHooTING y t i y t i . t i y t i y t i u l i u l i z i l error e21 is a data flash memory error and can be corrected by following the steps below: •...
  • Page 37: Exploded Views

    exploded views 220-240v model only 10x10 Hot Plate - Exploded View...
  • Page 38 XPloDeD IeWS 10x10 Hot Plate - Top Assembly...
  • Page 39 XPloDeD IeWS 220-240v model only 7x7 Hot Plate - Exploded View...
  • Page 40 XPloDeD IeWS 7x7 Hot Plate - Top Assembly...
  • Page 41: Wiring Diagram

    Wiring Diagram CONTROL T/C OTP T/C COMPONENT CHART REF. DESCRIPTION (7x7) (10x10) INLET CEX373 CEX373 EXT. PROBE CEX376 CEX376 RS232 WH1318X1 WH1318X1 EL1 (100V) ELEMENT ELX51 ELX54 EL1 (120V) ELEMENT ELX50 ELX53 EL1 (220-240V) ELEMENT ELX52 ELX55 F1 (220-240 ONLY) FUSE 261575 261575...
  • Page 42: Replacement Parts List

    replacement parts List To insure your safety and for proper operation, the ceramic top plates for hot plates and stir plates are only sold as complete assemblies. This assembly includes the ceramic top, element, thermocouple, insulation, baffle plate, and 2 ceramic top holders. We recommend not replacing individual components of the top plate.
  • Page 43: Accessories

    accessories Part No. Description TC732X1 General purpose immersion probe with 6” (15.24 cm) stainless steel sheath TC732X2 General purpose immersion probe with 10” (25.40 cm) stainless steel sheath TC727X2 Chemically-resistant immersion probe with 7” (17.78 cm) Teflon sheath 711S Non-Mercury Thermometer 20 to 100°C range - 76 mm immersion 647-1S Non-Mercury Thermometer 0 to 110°C range - 35 mm immersion 1007-3BLS...
  • Page 44: Ordering Procedures

    ordering procedures please refer to the Specification plate for the complete model number, serial number, and series number when requesting service, replacement parts or in any corre- spondence concerning this unit. all parts listed herein may be ordered from the Thermo Scientific dealer from whom you purchased this unit or can be obtained promptly from the factory.
  • Page 47: Warranty

    Two Year Limited Warranty This Thermo Scientific product is warranted to be free of defects in materials and workmanship for two (2) years from the first to occur of (i) the date the product is sold by the manufacturer or (ii) the date the product is purchased by the original retail customer (the “Commencement Date”).