5. Rotate the knob, select "View Step=1/6", and press [Enter] key for confirma-
6. Press [Esc] key, and exit the existing page.
Run List file
After the List files are edited, the user needs to enable the List function and re-
turn to the main interface to trigger the running.
1. Select the List file to be run in the Recall List.
2. In the List function setting interface, select "Run", and press [Enter] key to
3. Press [On/Off] to turn on the instrument output.
4. Conduct trigger based on the trigger mode set in the system menu, such as
Stop running and re-running
During the running of List file or when the running is finished, the interface dis-
plays END. Press [Shift]+[I-set](Function) composite key, and the instrument
interface displays the running state control interface of the existing file. The user
can press Stop to stop existing test and turn off the List function. Or, press
Reset to restart a new test.
The contents are consistent with the List Setup contents during editing of the
List file.
tion. Set the number of List step to be viewed, and press [Enter] key to enter
the parameter viewing page of the selected Step.
S001 Value=1.000V
S001 Slope=0.001S
S001 Time=0.005S
S002 Value=2.000V
S002 Slope=0.001S
S002 Time=0.005S
...
Rotate the knob for scrolled viewing of parameters of all Steps of the List.
The voltage and current parameters in the step vary with the existing List
mode.
confirm that the instrument automatically returns to the main interface and
waits for trigger.
0.000V
0.0000A
0.0W
WTG
front panel trigger mode. Press [Shift]+[On/Off] (Trigger) key for triggering,
and the instrument would output based on set sequence.
12.000V
5.0000A
100.00W
RUN
Copyright © Itech Electronic Co., Ltd.
Load Function
88