Wavelength Accuracy - Jasco V-530 Hardware/Function Manual

Uv/vis spectrophotometer
Hide thumbs Also See for V-530:
Table of Contents

Advertisement

(6) Measure the single beam spectrum with the sample chamber empty three
times to obtain three D
measurement example is shown in Fig 4.4.
(7) The wavelength repeatability is normal if it is within the following criterion.
Within ±0.1 nm of the average value of the peak of the emission
Criterion:
line spectrum at 656.1 nm
(8) Display the [Others] screen again and return to the original settings (WI and
D2).

4.4.3 Wavelength Accuracy

Measure the spectra of the D
nm and read the peak wavelength.
(1) Light only the Deuterium lamp by following items (1) to (3) in Section 4.4.2.
(2) Select [Single Beam] from the [Environment] menu to display the [Single
Beam] screen. (Refer to Section 9-5 of the Operations manual.)
(3) Display the [Parameter] screen and set the following parameters.
<<Parameters for spectrum with an emission line 656.1 nm>>
V-530 measurement parameters
Phtmtrc mode
Sam
Response
Fast
Start WL
658 nm
End WL
654 nm
Data interval
0.1 nm
Scan. speed
40 nm/min
No. of Cycles
1
Lower scale
0
Upper scale
100
Note:
PMT voltage
<<Parameters for spectrum with an emission line 486.0 nm>>
Start WL
End WL
All other parameter settings are the same as shown above.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
lamp spectra with an emission line at 656.1 nm. A
2
lamp with emission lines at 656.1 nm and at 486.0
2
V-550/560/570
parameters
Phtmtrc mode Sam
Response
Band width
Start WL
End WL
Data interval
Scan. speed
No. of Cycles
Lower scale
Upper scale
HT Voltage
with the instrument. Set accordingly.
varies
488 nm
484 nm
27
measurement
Fast
0.5 nm
658 nm
654 nm
0.025 nm
10 nm/min
1
0
100
300 V