Installation Conditions - Jasco V-530 Hardware/Function Manual

Uv/vis spectrophotometer
Hide thumbs Also See for V-530:
Table of Contents

Advertisement

Installation Conditions

To ensure operation safety, observe the following conditions:
(1)
Do not operate the instrument under voltage fluctuations exceeding 10%
of the recommended line voltage. Otherwise, the instrument may not
function properly.
(2)
Use a three-prong electrical outlet with a ground.
(3)
Operate the instrument in a temperature range of 10
(4)
Operate the instrument in a humidity range of 35%
humidity exceeds 85%(RH), condensation may deteriorate optical
components.
(5)
Operate the instrument in an atmospheric pressure range of 750
1060hPa.
(6)
Avoid strong magnetic fields and sources of high frequency. The
instrument may not function properly when near a strong magnetic field or
high frequency source.
(7)
Avoid vibration from vacuum pumps, electric motors, processing
equipment and machine tools.
(8)
Avoid dust and corrosive gas. Do not install the instrument where exposed
to dust, especially in locations exposed to outside air or ventilation outlets
that discharge dust particles.
(9)
Do not install the instrument where exposed to direct sunlight.
(10)
Ensure that no air conditioner blows air directly on the instrument. This
may inhibit stable measurement.
Note:
The above conditions do not ensure optimal performance of this
instrument.