1.1 V-530 UV/VIS Spectrophotometer Specifications

Optical system
Resolution
Light source
Light
source
changeover
wavelength
Wavelength range
Wavelength repeatability
Wavelength accuracy
Spectral bandwidth
Wavelength display
Photometric mode
Photometric range
Photometric display
Photometric repeatability
Photometric accuracy
Response
Number of integrations
Stray light
Wavelength scanning
Wavelength moving speed
Data pitch
Baseline stability
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Single monochromator
UV/VIS region: 1200 lines/mm concave grating
Rowland off-circle arrangement
Double beam type
2 nm
Deuterium lamp: 190 to 350 nm
Halogen lamp: 330 to 1100 nm
Any wavelength between 330 and 350 nm can
be selected.
190 to 1100 nm
±0.1 nm
±0.3 nm
2 nm fixed
In increments of 0.1 nm
Abs, %T. %R
-2.000 to 3.000 Abs (0 to 200 %T)
±10000 %T
-2 to 5 Abs
Four digits and (-)symbol
±0.001 Abs (0 to 0.5 Abs)
±0.002 Abs (0.5 to 1 Abs)
±0.002 Abs(0 to 0.5 Abs)
±0.004 Abs(0.5 to 1 Abs)
±0.3 %T
Note: Tested with NIST SRM 930D
Quick, Fast, Medium, Slow
2, 8, 32, 128 times (1 datum/30msec)
0.04 % (220 nm: NaI 10g/L aqueous solution;
340 nm: NaNO
50g/L aqueous solution)
2
40, 100, 200, 400, 1000, 2000, 4000 nm/min
8000 nm/min
0.1, 0.2, 0.5, 1, 2, 5, 10 nm/data
(spectrum measurement)
0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10 sec /data
(time course measurement)
±0.001 Abs/hour
1