4.5.2 Lamp Replacement and Adjusting the Light Source Mirror

The lifetime of both the Deuterium(D
1000 hours. After prolonged use, luminous energy decrease in turn causing the
level of noise in the measured data to increase.
4.5.2.1 Lamp Replacement
WARNINGS:
(1) The lamp is hot when lit. Therefore, before replacing the lamp,
switch off the light source and wait for at least 10 minutes.
(2) Before attempting to replace the lamp, turn off the power and
remove the power cable from the outlet.
CAUTIONS:
(1) Loosen only the screws required for lamp replacement.
(2) When handling the lamp, ware clean, cloth gloves. Never
handle the lamp with bare hands.
(3) Never touch the mirror or any other optical elements by hand.
(4) If the lamp surface is contaminated, clean it with gauze soaked
in ethanol and wipe it with dry clean gauze.
<<Replacement of Halogen(WI) lamp>>
(1) Slide the outer lid of the light source unit to remove it and then remove the
inner lid (Fig 4.8).