2. On the Measurements and Warning screen in the Run Group Configuration, set the "To Product"
value to 0.5 inch. This is an arbitrary beginning value and will be changed to the actual value later.
Note: Do not change the "To Cal Tiles" value.
3. Press the Control button on the toolbar to access the Sensor Control Dialog.
4. Press the Z Read button to move the z-axis to the Read position (fully-extended).
5. Manually measure the distance between the edge of the z-axis window bezel and the product. The
bezel is the lowest point on the z-axis assembly. You will need to either measure this distance in
inches or convert the measurement to inches (1 inch = 25.4 mm).
6. Subtract 1 inch (25.4 mm) from the value measured in #5.
7. On the Measurements and Warning screen in the Run Group Configuration, set the "To Product"
value to the number calculated in #6.
Note: Do not change the "To Cal Tiles" value.
8. Press the Control button on the toolbar to access the Sensor Control Dialog.
9. Press the Z Read button to move the z-axis to the Read position (fully-extended).
10. Repeat Steps 2-9 whenever product height changes. The physical position of the sensor will not
need to be changed unless the proper position cannot be achieved only through software
modifications (i.e., the z-axis is fully extended and is still too far from the product).