Setting channel coupling

To apply a signal to Channel A. The signal being
tested is a sine signal that contains DC quantities.
Press [A] to select Channel A. Then press [F1] to set
A input to "ON".
Next press [F2] to select DC coupling. Both DC and
AC quantities of the signal being tested and input to
Channel A can pass through.
Press [F2] to select AC coupling. DC quantities of
the signal being tested and input to Channel A will be
intercepted. Only AC quantities can pass through.
Press [F2] to select ground coupling. The channel
input is now set to equivalent ground. The channel
input level appears on screen.
Note: In this mode, although waveform is not displayed,
the input signal remains connected to the channel circuit.
Setting the channel bandwidth limit
Applying a signal to Channel B. The signal being
tested is a signal stacked with high frequency
quantities.
Press [B] to turn Channel B on. Next press [F3] to
set channel bandwidth to full bandwidth. The signal
being measured can pass through even if it contains
high frequency quantities.
Press [F3] to set bandwidth limit to 20MHz. All high
frequency quantities higher than 20MHz in the signal
being tested will be limited.
Probe ratio
1X
10X
100X
1000X
Polarity
Normal
Invert
Select one of the values based on the probe
attenuation factor to keep the reading of the
signal being measured correct.
Normal waveform display
Waveform is inverted
7