High Spo 2 Alarm; Low Spo - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.12 High SpO
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
This alarm requires a connected SpO

7.3.13 Low SpO

Property
Alarm text
Priority
Definition
Possible cause
Setting range
Setting resolution
Ventilator action
This alarm requires a connected SpO
Doc. 007229 Y-2
Alarm
2
Description
High SpO2
Medium
A High SpO
exceeds the alarm limit for 30 seconds.
Too high flow of bleed-in oxygen.
The SpO
The Vivo 2 will continue treatment according to the current
settings.
• 80 % to 100 %
• 90 % to 100 %
• Off
1 %
Alarm
2
Description
Low SpO2
High
A Low SpO
below the alarm limit for 30 seconds.
• Too low flow of bleed-in oxygen.
• Oxygen inlet is disconnected.
• Delivered tidal volumes are too small.
70% to 100%
1%
The ventilator will continue treatment with the same settings.
alarm will be given when the measured SpO
2
value goes back below the alarm limit.
2
sensor.
2
alarm will be given when the measured SpO
2
sensor.
2
2
is
2
Alarms 85