Low Pressure Alarm - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.2

Low Pressure Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Setting display
Doc. 007229 Y-2
Description
Low Pressure
High
A Low Pressure alarm will be given when the Vivo 2 pressure
fails to reach the low pressure alarm limit for 15 seconds.
• Disconnection of patient circuit.
• Mismatch between pressure setting and alarm setting.
• Leakage from the mask or other components of the
patient circuit.
The pressure rises above the alarm limit.
The Vivo 2 will continue treatment according to the current
settings.
• 1 cmH
O to 30 cmH
2
• Off
Note that the Low pressure alarm cannot be set higher than
the value set for IPAP or the High pressure alarm.
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
The Low Pressure alarm setting is displayed by a red line in
the pressure bar graph.
O
2
O
2
O
2
Vivo 2 User's Manual
Alarms 79