Alarm Message on the Screen
The name of the active alarm is displayed on the screen.
• High priority alarms
Red highlight color.
• Medium priority alarms
Yellow highlight color.
For detailed information about specific alarms, see 7.3 Physiological Alarms, page 78
and 7.4 Technical Alarms, page 87.
Display of Multiple Alarms
If several alarms are active, active high priority alarms have precedence over medium
priority alarms: All high priority alarm conditions must be resolved before any medium
alarms are displayed.
If several alarms of the same priority are active at the same time, the alarm descriptions
are looped in the display.
A ">>" symbol indicates that more alarms are to be displayed in the loop.
Alarm LED signal
• High priority alarms
Red light, flashing quickly (0.5 s.
interval).
• Medium priority alarms
Yellow light, flashing slowly (2 s.
interval).
7.2.2

Pause the Alarm Sound

The audible signal can be paused for 60
seconds by pressing the Audio Pause
button. Pressing the button again reac-
tivates the sound.
If a new alarm condition occurs during
the audio pause period, the audible sig-
nal will be reactivated.
Doc. 007229 Y-2
Alarms 75
Vivo 2 User's Manual