CAUTION!
Supplemental oxygen flow must not exceed 30 l/min or 100 kPa.
CAUTION!
Supplemental oxygen is added before the volume measurement sensor and thereby
included in the measurements. However, the oxygen concentration still has influence
on the volume measurement of the delivered air.
This measurement is based on a normal oxygen concentration of 21%. If the oxygen
concentration is higher, the actual inspired volume will deviate from the monitored
volume as follows:
• 40% oxygen concentration: -2.5% deviation
• 60% oxygen concentration: -5% deviation
• 80% oxygen concentration: -7.5% deviation
2.9
Mobile Use — Warning and Precautions
This section applies if using the Vivo 2 during transit, for example on a wheel chair or
in a car.
No Attachable Humidifier During Mobile Use
WARNING!
Risk of Suffocation
Do not use the attachable humidifier during mobile use.
Due to movements, water spillage from the humidifier or condensed water may flow
to the patient and cause suffocation.
WARNING!
Risk of Electric Chock
Do not use the attachable humidifier during mobile use.
Internal water spillage may cause electric chocks and may damage the device.
Doc. 007229 Y-2
Safety Information 25