Using The Oxygen Supply Adapter - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

CAUTION!
• When using the Vivo 2 with the SpO2 sensor, the Vivo 2
displays functional oxygen saturation measured by the sensor.
• The following information concerns the light emitted by the SpO2
– Peak Wavelength (red): 660 nm
– Peak Wavelength (infrared): 905 nm
– Maximum Optical Output Power: ≤ 15 mW
– For more information regarding the oxygen probe's range of peak wave-
lengths, max optical power and usage, please refer to the respective probe
manual.
• Environmental factors may influence the function or accuracy of the pulse oxi-
meter, such as ambient light, physical movement, diagnostic testing, low perfusion,
electromagnetic interference, dysfunctional haemoglobin, presence of certain dyes
and inappropriate positioning of the pulse oximeter probe.
• A functional test cannot be used to assess the accuracy of a pulse oximeter probe
or a pulse oximeter.
6.6.4

Using the Oxygen Supply Adapter

WARNING!
Read the section 2.8 Oxygen Usage — Warning and Precautions, page 24 before using the
Vivo 2 with oxygen.
Supplemental oxygen with a flow up to 30 l/min can be added by an oxygen source
with rotameter such as oxygen cylinder, central oxygen supply system or an oxygen
concentrator.
68 How to Use the Vivo 2
Vivo 2 User's Manual
Doc. 007229 Y-2